FDA Medical Device Recalls (openFDA)

American Medical Systems: AMS 700 CX MS Pump IZ, Preconnect Penoscrotal, 18 cm, Sterile EO, REF 72404232, Lot 551668001. Expiration Date 2010-Jun-20, Manufacture Date 2008 Jun 20.

Recall number
Z-2461-2008
Recalling firm
American Medical Systems
Product
AMS 700 CX MS Pump IZ, Preconnect Penoscrotal, 18 cm, Sterile EO, REF 72404232, Lot 551668001. Expiration Date 2010-Jun-20, Manufacture Date 2008 Jun 20.
Status
Terminated
Initiated
2008-07-17
Posted
2008-09-20
Terminated
2012-05-01
Reason
The carton labeling does not match the product packaged inside. The package indicated that the device is an 18cm Preconnected MS pump while it is actually an 18cm Ultrex Preconnected MS Pump.
Root cause
Employee error
Action
On July 17, 2008, the physician who implanted the device received a phone call from AMS to discuss the event and be sure he understands what went wrong (Ultrex cylinders implanted instead of CX) and any new risks introduced to his patient so he can properly assess any impact it may have. A follow up letter " AMS Important Product Recall" dated July 21, 2008 was also sent to document the event and the call. The letter requested to return of the "Recall Acknowledgement Form". The second physician (tried not used) has already returned the device and will not receive further communication. A complaint was assigned to document the event.
Quantity
5 units
Distribution
LA
Lot, serial or model codes
Lot # 551668
510(k) numbers
not on file
PMA numbers
not on file
Product code
JCW
Firm FEI
1000116179
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register