Skip to the content

FDA Medical Device Recalls (openFDA)

Cholestech Corp: Cholestech GDX A1C Test Cartridge. The Cholestech GDX A1C Test is an in-vitro diagnostic test that measures hemoglobin A1c (HbA1c).

Recall number
Z-2460-2008
Recalling firm
Cholestech Corp
Product
Cholestech GDX A1C Test Cartridge. The Cholestech GDX A1C Test is an in-vitro diagnostic test that measures hemoglobin A1c (HbA1c).
Status
Terminated
Initiated
2006-08-03
Posted
2008-09-23
Terminated
2009-10-07
Reason
Cholestech Corp. was notified by the foreign manufacturer of the A1C test cartridge that stability data indicated that the product would not meet performance claims through the end of its shelf life.
Root cause
Nonconforming Material/Component
Action
A customer notification letter was sent to Cholestech consignees who had received product from this lot. The letter dated 8/3/2006 instructed customers to discontinue use of the lot of product, destroy product according to in-house standard procedures and contact Cholestech Technical Services for replacement. It also instructed to provide a copy of the letter to other users if the lot of product was further distributed. A response form was included with instructions to complete and return. Contact Cholestech Corp at 1-510-781-3505 for assistance.
Quantity
9,324 Cartridges
Distribution
U.S. state of Colorado. OUS: India, Slovenia, Germany, Tunisia, Taiwan, Canada, Philippines, South Africa and Mexico
Lot, serial or model codes
Catalog #12-658, Lot #066T28
510(k) numbers
["K041635"]
PMA numbers
not on file
Product code
LCP
Firm FEI
1000123985
Firm city
Hayward
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register