Skip to the content

FDA Medical Device Recalls (openFDA)

Ethicon, Inc.: PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067

Recall number
Z-2459-2018
Recalling firm
Ethicon, Inc.
Product
PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
Status
Terminated
Initiated
2018-05-07
Posted
not on file
Terminated
2020-06-08
Reason
The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.
Root cause
Package design/selection
Action
On May 24, 2018, the firm distributed Urgent: Medical Device Recall letters to affected U.S. customers. Customers were instructed to do the following: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal) on hand and quarantine such product(s). 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. 4. Complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-888-473-8015 or Ethicon3681@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall (removal). 5. Keep this notice visibly posted for awareness until all product subject to this recall (removal) has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. 6. Customers are required to return all unused D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture products that are in their inventory immediately. Only D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture product subject to this recall (removal) returned by September 21, 2018 will be eligible for credit reimbursement. Any product subject to this recall (removal) returned after September 21, 2018 will not be eligible for credit reimbursement. 7. To return product subject to this recall (removal), photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for
Quantity
1020
Distribution
Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
Lot, serial or model codes
KKJ659 LAH044 LEH699 MCH331 MDH290 MCQ536
510(k) numbers
["K133356"]
PMA numbers
not on file
Product code
GAW
Firm FEI
2210968
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register