Skip to the content

FDA Medical Device Recalls (openFDA)

AMD Medicom Inc.: Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C

Recall number
Z-2458-2026
Recalling firm
AMD Medicom Inc.
Product
Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C
Status
Open, Classified
Initiated
2026-05-08
Posted
2026-06-16
Terminated
not on file
Reason
Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Root cause
Material/Component Contamination
Action
On 5/8/2026, recall notices were emailed to customers who were asked to do the following: 1) Discontinue use, quarantine and destroy affected devices. 2) Notify any customers or facilities to whom you have distributed this product. 3) Complete and return the response form via email to firm's customer service at Claims@medicom.ca On 6/5/2026, recall notices with all affected lots were emailed to customers.
Quantity
7,838,200
Distribution
US: MT, NC, VA, CA, TX, OH, TN, IL, RI
Lot, serial or model codes
UDI-DI: 20686864043205. Distributed After May 1st, 2022, including the following lots: 19316, 16409, 15463-7, 15463-6, 15463-5, 15463-4, 15463-3, 15463-2, 15463-1, 15463, 15192-1, 15192, 14813, 14039, 13673, 13289, 12641, 12206, 11851, 11020-1, 11020, 10586, 10392, 9243, 8160, 7784, 7748, 7044
510(k) numbers
["K830269"]
PMA numbers
not on file
Product code
EFQ
Firm FEI
3002677549
Firm city
Montreal
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register