Skip to the content

FDA Medical Device Recalls (openFDA)

CAREFUSION: BD BURETTE GRAVITY SET 60 DROP 2 Smartsite VALVES VENTED/NONVENTED, REF: 42163E

Recall number
Z-2458-2021
Recalling firm
CAREFUSION
Product
BD BURETTE GRAVITY SET 60 DROP 2 Smartsite VALVES VENTED/NONVENTED, REF: 42163E
Status
Open, Classified
Initiated
2021-08-02
Posted
not on file
Terminated
not on file
Reason
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Root cause
Process control
Action
On 08/02/21, Urgent Medical Device Correction notices were mailed and emailed to consignees who were asked to notify recalling firm if this issue was experienced at 1-844-823-5433, productcomplaints@bd.com . Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device. If during priming the clinician encounters an occluded needle-free connector, take the following actions: 1) Depress the syringe plunger while the syringe is still attached to the needle-free connector. 2) If unable to depress the plunger, pull back on the syringe plunger and depress the syringe plunger again. Repeating this step may assist in opening the needle-free connector valve. 3) If the valve still is not open; without disconnecting the syringe male luer tip from the BD SmartSite valve surface, carefully unthread and rethread the syringe to partially disengage and reengage the syringe. Repeating this step up to three times may assist in opening the needle-free valve. Note: To maintain sterility, do not allow the syringe and needle-free valve to completely disconnect. 4) If previous steps do not open the valve, then open a new needle-free connector to replace the occluded connector. 5) After the steps above, proceed with the existing instructions for use. 6) Share this correction notification with all users within your facility network of the product to ensure they are also aware of this correction. 7) Complete and return the Customer Response Form. Distributors were asked to identify customers within their distribution network that purchased the affected product and to email an excel file listing of all customers to the recalling firm within 72 hours of receipt of this letter so the recalling firm can initiate customer notification. See online tool to support the identification of impacted lot numbers https://www.bd.com/MDS-21-4144
Quantity
630
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR.
Lot, serial or model codes
Lots: 21015553
510(k) numbers
["K820278"]
PMA numbers
not on file
Product code
FPA
Firm FEI
3012410077
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register