Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries Inc: Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Weeks Kit (DYNDC1987A) is intended for use on patients greater than 34 weeks of age. The central venous catheter, also called a central line, is a long, thin, flexible tube used to give medicines, fluids, nutrients, or blood products over a long period of time, usually several weeks or more. A catheter is often inserted in the arm or chest through the skin into a large vein. The catheter is threaded through this vein until it reaches a large vein near the heart.

Recall number
Z-2458-2015
Recalling firm
Medline Industries Inc
Product
Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Weeks Kit (DYNDC1987A) is intended for use on patients greater than 34 weeks of age. The central venous catheter, also called a central line, is a long, thin, flexible tube used to give medicines, fluids, nutrients, or blood products over a long period of time, usually several weeks or more. A catheter is often inserted in the arm or chest through the skin into a large vein. The catheter is threaded through this vein until it reaches a large vein near the heart.
Status
Terminated
Initiated
2015-08-03
Posted
2015-08-25
Terminated
2017-01-05
Reason
The print plate artwork for this lot was found to state: Central Line <34 Weeks Kit displaying an incorrect sign in front of the 34. The description for this device should have stated Central Line >34 Weeks Kit. The defect may lead to the product to be used on patient population that is <34 weeks rather than intended population of >34 weeks.
Root cause
Error in labeling
Action
Medline Industries contacted the account directly via phone on August 4, 2015, they confirmed no affected product is in their facility. The lot numbers are easily identified on the outside of each kit. Second and third follow up communications will not be sent as we have confirmed the facility does not have any affected product. For questions regarding this recall call 847-643-4129.
Quantity
320 units for lot 14DB9277 & 320 units for lot 14PB2878
Distribution
US including PA
Lot, serial or model codes
Central Venous Catheter Kit for patients >34 weeks old with the coding below were affected.   Lot Numbers: 14DB9277 and 14PB2878 Unit No: DYNDC1987A Expiration Date: 06/2015  Serial No: N/A UPC code: N/A
510(k) numbers
not on file
PMA numbers
not on file
Product code
OFF
Firm FEI
1417592
Firm city
Mundelein
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register