Conmed Linvatec Endoscopy Division: ConMed Linvatec Surgical Video Cart, Catalogue Number: VP8500

Recall number
Z-2458-2008
Recalling firm
Conmed Linvatec Endoscopy Division
Product
ConMed Linvatec Surgical Video Cart, Catalogue Number: VP8500
Status
Terminated
Initiated
2007-10-31
Posted
2008-09-21
Terminated
2008-09-22
Reason
The recall was initiated because ConMed Linvatec has determined there is a possibility the wheel caster(s) may become loose which may result in the wheel caster(s) falling out of the base of the VP8500 Video Cart. The potential for the wheel caster or casters to fall out of the base of the cart may result in a potential tip hazard.
Root cause
Device Design
Action
The recall was initiated by the firm sending the recall information packet via UPS overnight delivery to the Distribution and Sales Force in May 2008. The packet included: 1. Recall Notification and reply form for each consignee. 2. Instructions on the procedure to conduct the recall of the VP8500. 3. A list of affected consignees in their territory. The Recall Notification informed the customers of the potential for the wheel caster(s) to fall out of the base of the cart which may result in a potential tip hazard. The customers were instructed to discontinue the use of the VP8500 Video Cart immediately if the wheel caster(s) show any signs of being loose and to complete the recall reply form. Upon receipt of notification, distributors are supposed to : 1. Visit the affected accounts and provide them with a copy of the Recall Notification to inform them of the recall and the Recall Acknowledgment Reply Form which is to be mailed or faxed back to the firm. 2. Schedule the delivery and installation of refurbished carts to replace the recalled cart(s) by contacting Jill Cover, ConMed Linvatec Endoscopy Division at (888) 292-0100, ext. 8157 or e-mail jcover@livatec.com to arrange for replacement. 3. Move medical devices, etc. from the recalled cart(s) and set up the refurbished cart(s) with the required equipment at the account's location. (ProMEDICA will arrange for shipping of the recalled cart(s) back to ProMEDICA.)
Quantity
45 Units Worlwide (42 in US)
Distribution
Worldwide Distribution including states of: AK, AZ, CA, FL, GA, IL, IN, KS, MD, NC, NY, OH, and PA and country of Canada.
Lot, serial or model codes
Catalog Number: VP8500;   Serial #;s: 6080185, 6080285, 6080385, 6080485, 6080585, 6080685, 6080785, 6080885, 6080985, 6081085, 6081185, 6081285, 6081385, 6081485, 6081585, 6081685, 6081785, 6081885, 6081985, 6082085, 6082185, 6082285, 6082385, 6082485, 6082585, 6082685, 6082785, 6082885, 6082985, 6083085, 6083185, 6083285, 6083385, 6083485, 6083585, 6083685, 6083785, 6083885, 6083985, 6084085, 6090185, 6090285, 6090385, 6090485, and 6090585.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KQM
Firm FEI
1000125338
Firm city
Goleta
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register