Skip to the content

FDA Medical Device Recalls (openFDA)

Cholestech Corp: Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum. Measurement of CRP is useful as an aid in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. The test cassettes are for use with the LDX analyzer.

Recall number
Z-2457-2008
Recalling firm
Cholestech Corp
Product
Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum. Measurement of CRP is useful as an aid in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. The test cassettes are for use with the LDX analyzer.
Status
Terminated
Initiated
2006-08-09
Posted
2008-09-20
Terminated
2009-09-30
Reason
Test results are high, outside of control material upper limit specification. Results could be as high as 15%.
Root cause
Device Design
Action
Cholestech provided its three distributors a notification letter/response form with instructions to distribute the letter to their customers/end users dated August 9, 2006. The letter instructed customers to discontinue use of the lot of cassettes, return remaining cassettes to Cholestech through the Technical Service Representative and to complete and return the response form. The end users returned the signed letters directly to Cholestech so that effectiveness checks could be completed.
Quantity
2, 910 cassettes
Distribution
Nationwide Distribution --- including states of Virginia, Texas and North Carolina, Ohio, Arkansas, Washington, Florida , California and Michigan.
Lot, serial or model codes
Catalog #12-807, Lot #C17-6434
510(k) numbers
["K040579"]
PMA numbers
not on file
Product code
DCK
Firm FEI
1000123985
Firm city
Hayward
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register