FDA Medical Device Recalls (openFDA)
Smith & Nephew, Inc.: TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Recall number
- Z-2456-2026
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Status
- Open, Classified
- Initiated
- 2026-03-16
- Posted
- 2026-06-16
- Terminated
- not on file
- Reason
- Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.
- Root cause
- Process control
- Action
- On 3/16/2026, recall notices were emailed to customers who were asked to do the following: 1) Locate and quarantine affected devices immediately. If you have further distributed the product to other organizations, please inform them at once of this Field Action and provide to them a copy of this letter. 3. Please complete the Customer Response form and email or fax it to your national Smith+Nephew agency/distributor. 4. Return quarantined product to your national Smith+Nephew agency/distributor. 5. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. Firm contact information: + 1 901 396 2121, 1 800 821 5700
- Quantity
- 5
- Distribution
- International distribution to the countries of BE, DE, IT.
- Lot, serial or model codes
- UDI-DI: 03596010514394. Lot/Batch: 25BM03616
- 510(k) numbers
- ["K210980"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 1020279
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28