Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Recall number
Z-2456-2026
Recalling firm
Smith & Nephew, Inc.
Product
TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
Status
Open, Classified
Initiated
2026-03-16
Posted
2026-06-16
Terminated
not on file
Reason
Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.
Root cause
Process control
Action
On 3/16/2026, recall notices were emailed to customers who were asked to do the following: 1) Locate and quarantine affected devices immediately. If you have further distributed the product to other organizations, please inform them at once of this Field Action and provide to them a copy of this letter. 3. Please complete the Customer Response form and email or fax it to your national Smith+Nephew agency/distributor. 4. Return quarantined product to your national Smith+Nephew agency/distributor. 5. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. Firm contact information: + 1 901 396 2121, 1 800 821 5700
Quantity
5
Distribution
International distribution to the countries of BE, DE, IT.
Lot, serial or model codes
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
510(k) numbers
["K210980"]
PMA numbers
not on file
Product code
HSB
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register