Skip to the content

FDA Medical Device Recalls (openFDA)

LivaNova USA Inc.: Sterile Disposable Connectors ASY RDC Y 1/2X3/8X3/8 STR 24 REF/ GTIN for Insert Label:050529000/ 00803622123351

Recall number
Z-2456-2020
Recalling firm
LivaNova USA Inc.
Product
Sterile Disposable Connectors ASY RDC Y 1/2X3/8X3/8 STR 24 REF/ GTIN for Insert Label:050529000/ 00803622123351
Status
Terminated
Initiated
2020-04-24
Posted
not on file
Terminated
2022-02-17
Reason
The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
Root cause
Packaging process control
Action
The firm sent an URGENT MEDICAL DEVICE REMOVAL letter to affected customers on 04/24/2020. The purpose of this medical device correction is to: 1. Advise the customers of the present issue and not to use the affected devices; 2. Providing information to the user on product disposal or return and replacement; 3. Inform affected customers that LivaNova will coordinate replacement for the unused products.
Quantity
768 units
Distribution
US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)
Lot, serial or model codes
LOT 2003400086
510(k) numbers
["K833322"]
PMA numbers
not on file
Product code
DTL
Firm FEI
1718850
Firm city
Arvada
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register