Trilliant Surgical Ltd.: Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.

Recall number
Z-2456-2015
Recalling firm
Trilliant Surgical Ltd.
Product
Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.
Status
Terminated
Initiated
2015-07-20
Posted
2015-08-24
Terminated
2017-01-26
Reason
The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.
Root cause
Component design/selection
Action
The recalling firm sent out written notification of the recall on 7/3/14. The letter instructed the consignees to exchange affected instruments.
Quantity
638 units
Distribution
Nationwide throughout the US
Lot, serial or model codes
TSL000145, TSL000162, TSL000699, TSL000146, TSL000163, TSL000700, TSL00147, TSL000164, TSL000701, TSL000702, TSL000148, TSL000165, TSL000706, TSL000149, TSL000166, TSL000688, TSL000150, TSL000167, TSL000707, TSL000708
510(k) numbers
["K121452"]
PMA numbers
not on file
Product code
HTO
Firm FEI
3007420745
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register