Skip to the content

FDA Medical Device Recalls (openFDA)

mo-Vis BVBA: Micro Joystick R-net. Electrical wheelchair component.

Recall number
Z-2454-2025
Recalling firm
mo-Vis BVBA
Product
Micro Joystick R-net. Electrical wheelchair component.
Status
Open, Classified
Initiated
2025-07-10
Posted
2025-09-02
Terminated
not on file
Reason
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Root cause
Device Design
Action
An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing contact@mo-vis.com to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to contact@mo-vis.com or by phone at +32 9 335 28 60.
Quantity
3 units
Distribution
US Nationwide distribution in the state of TX.
Lot, serial or model codes
Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KNN
Firm FEI
3011795161
Firm city
Nevele
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register