Skip to the content

FDA Medical Device Recalls (openFDA)

Becton, Dickinson and Company, BD Biosciences: BD CD56 PE Cy 7: NCAM16.2 ASR Analyte Specific Reagent, Catalog number 335809 Hematology: The CD 56 antigen promotes homophilic adhesion in neurons and may be involved in homophilic adhesion of NK cells.

Recall number
Z-2453-2015
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD CD56 PE Cy 7: NCAM16.2 ASR Analyte Specific Reagent, Catalog number 335809 Hematology: The CD 56 antigen promotes homophilic adhesion in neurons and may be involved in homophilic adhesion of NK cells.
Status
Terminated
Initiated
2015-07-17
Posted
2015-08-24
Terminated
2016-03-11
Reason
Three lots of CD56 PE Cy 7, Catalog number 335809, contain a low amount of HLA-DR PE antibody which could result in an unexpected staining pattern.
Root cause
Material/Component Contamination
Action
BD Bioscience sent an Urgent Product Recall Letter dated July 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer Actions : 1. If you have CD56 PE-Cy7 integrated into a Laboratory Developed Test (LDT) or panel, evaluate the impact of a false positive in the PE channel the interpretation of on the results. 2. Discard all remaining stock of CD56 PE-Cy7, Catalog Number 335809, lot number 4300708, 4304864, and/or 5013643, and request replacement product. BD will replace product at no cost. 3. As a requirement of our Regulatory Tracking Process, we request that you complete the enclosed Tracking/Verification form, even if you no longer have the affected reagents listed in your inventory. Please return the Tracking/ Verification form to BD Biosciences, per the instructions (mail, fax, or email). We apologize for any inconvenience that this may have caused and are working toward a long-term solution to this issue. If you require further assistance, please contact BD Customer Support at 877.232.8995 (prompt 3) in the United States. For customers outside the US, contact your local BD Biosciences representative or distributor.
Quantity
362 units
Distribution
US Distribution and Internationally to Taiwan.
Lot, serial or model codes
Catalog number 335809; Lot numbers 4300708, 4304864, 5013643; all with expiry dates of 31 OCT 2015
510(k) numbers
not on file
PMA numbers
not on file
Product code
MVU
Firm FEI
2916837
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register