FDA Medical Device Recalls (openFDA)
LivaNova USA Inc.: Sterile Disposable Connectors ASY CONN REDUCER 1/2 X 3/8 24 REF/ GTIN for Insert label:050518000/ 00803622123283
- Recall number
- Z-2452-2020
- Recalling firm
- LivaNova USA Inc.
- Product
- Sterile Disposable Connectors ASY CONN REDUCER 1/2 X 3/8 24 REF/ GTIN for Insert label:050518000/ 00803622123283
- Status
- Terminated
- Initiated
- 2020-04-24
- Posted
- not on file
- Terminated
- 2022-02-17
- Reason
- The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
- Root cause
- Packaging process control
- Action
- The firm sent an URGENT MEDICAL DEVICE REMOVAL letter to affected customers on 04/24/2020. The purpose of this medical device correction is to: 1. Advise the customers of the present issue and not to use the affected devices; 2. Providing information to the user on product disposal or return and replacement; 3. Inform affected customers that LivaNova will coordinate replacement for the unused products.
- Quantity
- 96 units
- Distribution
- US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)
- Lot, serial or model codes
- LOT 2003400083
- 510(k) numbers
- ["K833322"]
- PMA numbers
- not on file
- Product code
- DTL
- Firm FEI
- 1718850
- Firm city
- Arvada
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28