Siemens Healthcare Diagnostics, Inc.: Dimension Vista B2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Recall number
Z-2450-2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Vista B2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.
Status
Terminated
Initiated
2015-07-23
Posted
2015-08-22
Terminated
2017-03-24
Reason
customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.
Root cause
Environmental control
Action
Siemens sent an Urgent Medical Device Correction letter dated July 2015 to all affected customers. The letter identified the affected product, problem, risk to health, and actions to be taken by customers. For questions contact your Customer Care Center - Technical Solutions at 1-800-441-9250 or your local Siemens technical support representative.
Quantity
Total 1655 units
Distribution
US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI
Lot, serial or model codes
Lot Number:15037MA; ExpDate: 2016-07-14 Lot # 15175MA; ExpDate: 2015-11-17; Lot # 15204MA; Expdate: 2017-01-13;  Lot #15246MA; expdate: 2017-02-26 Lot # 15267MA; expdate: 2017-02-26
510(k) numbers
["K083463"]
PMA numbers
not on file
Product code
JZG
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register