Carestream Health Inc: CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.
- Recall number
- Z-2449-2015
- Recalling firm
- Carestream Health Inc
- Product
- CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.
- Status
- Terminated
- Initiated
- 2015-08-07
- Posted
- 2015-08-22
- Terminated
- 2016-07-06
- Reason
- Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.
- Root cause
- Software design
- Action
- Carestream sent an Urgent Medical Device Correction letters dated July 29, 2015 and Customer Notification Acknowledgement Forms to customers via Certified Mail. The letter identified the affected product, problem and actions to be taken. Carestream will update the software on your device to correct this issue. For questions or concerns, contact the Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis. Outside of the U.S., please call your local Service support number.
- Quantity
- Domestic: 76 units, Foreign: 160 units
- Distribution
- Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.
- Lot, serial or model codes
- Software versions 5.6 and greater and 5.7 (current releases); Serial Numbers: 5029, 5569, 4067, 4754, 4131, 4100, 4096, 4076, 4569, 4570, 4588, 4670, 4767, 4879, 5266, 5393, 5408, 5442, 5572, 5573, 5555, 5556, 5563, 4146, 4679, 4702, 4703, 4704, 4801, 5090, 5154, 4787, 4069, 4074, 4344, 4085, 4058, 4084, 5470, 5158, 4086, 4099, 4073, 4665, 5241, 5303, 5304, 4639, 5350, 5352, 5429, 5508, 4696, 4763, 5182, 4291, 5037, 5554, 4192, 4117, 4172, 4207, 4225, 4239, 4278, 4301, 4310, 4328, 4333, 4338, 4366, 4384, 4421, 4427, 4442, 4443, 4447, 4465, 4499, 4523, 4534, 4538, 4539, 4692, 4822, 4861, 4875, 5128, 5131, 5174, 5553, 5579, 5580, 5597, 5628, 4320, 5309, 5042, 4807, 4808, 4809, 4810, 5493, 4266, 4711, 5327, 4125, 4143, 4097, 4156, 4282, 4287, 4289, 4258, 4742, 4964, 4784, 5050, 5551, 4233, 4644, 4082, 4113, 4196, 4683, 4309, 4315, 4064, 4256, 4987, 4624, 4585, 4882, 5362, 4004, 5377, 5418, 4300, 4308, 4627, 4638, 5138, 5251, 5360, 4947, 5620, 4985, 5403, 5495, 5560, 5544, 5545, 5621, 5622, 5451, 5282, 5645, 4724, 4298, 4269, 4756, 4418, 5273, 5178, 5179, 5443, 5197, 5389, 5584, 5473, 4899, 5334, 5431, 4990 --- NOTE: Additional serial numbers at foreign consignees to be determined upon reporting part usage.
- 510(k) numbers
- ["K141837"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 1315356
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28