Philips Medical Systems North America Co. Phillips: BIPLANE X-ray Imaging Systems included Allura XPER FD10/10, FD20/10, FD20/20, Integris BH5000, and Integris Allura 12-15 Biplane. The BIPLANE system has two C-arms and it allows the operator to image in two planes: Lateral and Frontal. It has a C-arm on the stand and C-arm suspended from the ceiling. The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands".
- Recall number
- Z-2449-2010
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Product
- BIPLANE X-ray Imaging Systems included Allura XPER FD10/10, FD20/10, FD20/20, Integris BH5000, and Integris Allura 12-15 Biplane. The BIPLANE system has two C-arms and it allows the operator to image in two planes: Lateral and Frontal. It has a C-arm on the stand and C-arm suspended from the ceiling. The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands".
- Status
- Terminated
- Initiated
- 2010-02-04
- Posted
- 2010-09-22
- Terminated
- 2011-09-23
- Reason
- A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
- Root cause
- Software design
- Action
- On 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator. This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm or cold system restart. A warm restart would take about 90 seconds or a cold restart would take about four minutes, and the device could not be used during this time. If the system does not work after the restart, the consignees should contact the Philips Healthcare Customer Service Representative at 1-800-772-9377 #5 (then follow prompts). A service representative will visit each consignee and conduct a software and hardware upgrade.
- Quantity
- 267
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Each unit is identified with a SITE number: 76992, 103526, 103581, 103921, 104020, 104021, 104068, 104386, 104497, 104628, 104926, 104950, 104953, 104959, 105184, 105399, 105581, 105585, 105654, 504655, 504674, 504733, 504742, 504779, 504868, 504921, 505067, 505068, 505120, 505356, 505380, 505530, 505611, 505703, 505903, 506008, 506026, 506069, 506119, 506509, 506655, 519118, 519136, 519196, 519197, 519678, 519730, 519981, 520275, 520582, 520666, 520688, 520756, 520793, 520949, 520964, 521149, 521325, 521525, 522136, 522456, 530396, 530653, 530654, 530660, 531649, 531758, 531799, 531926, 531939, 532152, 532218, 532240, 532241, 532383, 532614, 533288, 533290, 533397, 534131, 534156, 534786, 535345, 535761, 535775, 536028, 536031, 536034, 536035, 536037, 536042, 536171, 536715, 536789, 536899, 537175, 537593, 537631, 537804, 538028, 538095, 538099, 538157, 538172, 538387, 538409, 538514, 538516, 539213, 539326, 539628, 540184, 540237, 540461, 540721, 541237, 541238, 541239, 541276, 541577, 541589, 541590, 542382, 542413, 542557, 542714, 542963, 543289, 543411, 543690, 543879, 543953, 543989, 544306, 544395, 544543, 544635, 544656, 544668, 544844, 545025, 545262, 545396, 545518, 545583, 545604, 545753, 545882, 545884, 545997, 546013, 546119, 546149, 546569, 546906, 547173, 547215, 547431, 547642, 547715, 547769, 547823, 547922, 548005, 548020, 548036, 548082, 548140, 548288, 548383, 548618, 549202, 549242, 549368, 549370, 549903, 550198, 550691, 550758, 550812, 550935, 551855, 552355, 552599, 552639, 552969, 553685, 553699, 553766, 554763, 554802, 554970, 555293, 555299, 555356, 555834, 558604, 558945, 41443586, 41443609, 41443627, 41443647, 41443727, 41443765, 41443804, 41443909, 41444645, 41444647, 41444650, 41444689, 41444752, 41444816, 41445118, 41445432, 41445448, 41445501, 41445538, 41445615, 41445628, 41446230, 41446262, 41446773, 41455885, 41455897, 41455920, 41455925, 41455933, 41455944, 41455953, 41660520, 41780160, 41793290, 41795344, 41796154, 42046093, 42046440, 42057324, 42064058, 42067824, 42394072, 42479274, 42508759, 42553772, 42564341, 42625859, 42735090, 42891020, 42891095, 42891704, 43043666, 43058934, 43164789, 43173591, 43315510, 43436747, 43443622, 43453735, 43635004, 43670742, 43848923, 44028554, 44066424, 44584972, 44616211, 44654768, 44735751, and 44892370.
- 510(k) numbers
- ["K031333","K033737","K984545"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1000524572
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28