FDA Medical Device Recalls (openFDA)
Randox Laboratories Ltd.: Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691
- Recall number
- Z-2448-2023
- Recalling firm
- Randox Laboratories Ltd.
- Product
- Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691
- Status
- Open, Classified
- Initiated
- 2023-07-19
- Posted
- 2023-08-23
- Terminated
- not on file
- Reason
- As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.
- Root cause
- Under Investigation by firm
- Action
- Randox Labs Ltd (Manufacturer) extended recall to the US distributor via Medical Device Correction Letter dated July 19, 2023. The distributor initial contact with customer and follow up contacts via email and telephone. Letter states reason for recall, health risk and action to take: " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns please contact Rand ox Technical Services. Any customers using batches 590858 and 590859 for Ferritin, please contact technical.services@randox.com
- Quantity
- 166 units
- Distribution
- CA, FL, IL, ME, MI, NC, OH, PA,NJ,
- Lot, serial or model codes
- GTIN: 05055273204032 Batch/ Lot number Expiry Date 590858 28 08 23; 590859 28 08 23; 627222 28Jul 24; 627224 28Jul24; 634886 28Jul24; 634887 28Jul24
- 510(k) numbers
- ["K061056"]
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28