FDA Medical Device Recalls (openFDA)
Alcon Research, Ltd.: LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor Stand is an accessory to the LuxOR Opthalmic Microscope and the LuxOR Opthalmic Microscope with Q-VUE 3-D Assistant Visualization and provides a portable suspension system for the microscope that integrates an AC power module and microscope controls (Primary Control Panel and connectivity of foot controllers)
- Recall number
- Z-2448-2015
- Recalling firm
- Alcon Research, Ltd.
- Product
- LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor Stand is an accessory to the LuxOR Opthalmic Microscope and the LuxOR Opthalmic Microscope with Q-VUE 3-D Assistant Visualization and provides a portable suspension system for the microscope that integrates an AC power module and microscope controls (Primary Control Panel and connectivity of foot controllers)
- Status
- Terminated
- Initiated
- 2015-08-11
- Posted
- 2015-08-21
- Terminated
- 2016-11-04
- Reason
- Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.
- Root cause
- Device Design
- Action
- Alcon sent an Urgent Medical Device Correction letter dated August 11, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Alcon at 800-862-5266.
- Quantity
- 688 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, France, Germany, Greece, Guatemala, Hungary, India, Italy, Japan, Jordan, Korea, Lebanon, Malaysia, Mexico, Mongolia, New Zealand/Fiji, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, Ukraine and United Kingdom.
- Lot, serial or model codes
- All units LX3 Microscope Floor Stands manufactured prior to July 2015 (Serial Numbers 1402548501X to 1502857606X)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HRM
- Firm FEI
- 1610287
- Firm city
- Fort Worth
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28