FDA Medical Device Recalls (openFDA)
Customed, Inc: LAPAROSCOPY CHOLE SYSTEM PACK MIAMI METRO CON TENTS: (12) TOWELS CLOTH HUCK BLUE LIF (1) MAYO STAND COVER REINFORCED LIF (1) TABLE COVER REINFORCED 50" X 90" LIF (1) SUTURE BAG FLORAL LIF (10) GAUZE SPONGE 4" X 4" 16PLY XRD LIF (1) BULB SYRINGE 60cc LIF (1) BLADE SURGICAL #10 STAINLESS STEEL (2) SHEET% 60" X 77" DRAPE REINFORCED LIF (10) LAP SPONGE PREWASH 18" X 18" XRD LIF (1) BLADE SURGICAL #15 CARBON STEEL (1) TUBE SUCTION CONNECT Y." X 12' LIF (1) BAG GLASSINE PLAIN (1) CAUTERY TIP POLISHER LIF (1) DRAPE LAP. ABDOMINAL WITH POUCH 102" X 122" X 78" LIF (1) PK. STRIP STERI CLOSURE W' X 4" LIF (1) SKIN MARKER INK WITH 8/LABEL/TIME OUT/RULER L/F (1) CORD LAPAROSCOPY MONOPOLAR FOOD SWICHING (1) COUNTER NEEDLE BLADE 30c MAG/FOAM (1) TIME OUT BEACON NON WOVEN LIF (1) GOWN XL SMS IMPERVIOUS REINFORCED LIF (1) GOWN SURG. REINFORCED X LARGE TOWEL/WRAP LIF (1) TOWEL ABSORBENT 15" X 20" LIF (1) BLADE SURGICAL #11 CARBON STEEL (1) TUBING INSUFFLATION OPEN TIP WITHOUT VENT LIF (1) YANKAUER OPEN TIP CLEAR WITHOUT C IV LIF ATTACHED PACK WITH THE FOLLOWING STERILE COMPONENT: (1) ACTIFOG WITH/FOAM PAD SOLUTION LIF Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Recall number
- Z-2447-2014
- Recalling firm
- Customed, Inc
- Product
- LAPAROSCOPY CHOLE SYSTEM PACK MIAMI METRO CON TENTS: (12) TOWELS CLOTH HUCK BLUE LIF (1) MAYO STAND COVER REINFORCED LIF (1) TABLE COVER REINFORCED 50" X 90" LIF (1) SUTURE BAG FLORAL LIF (10) GAUZE SPONGE 4" X 4" 16PLY XRD LIF (1) BULB SYRINGE 60cc LIF (1) BLADE SURGICAL #10 STAINLESS STEEL (2) SHEET% 60" X 77" DRAPE REINFORCED LIF (10) LAP SPONGE PREWASH 18" X 18" XRD LIF (1) BLADE SURGICAL #15 CARBON STEEL (1) TUBE SUCTION CONNECT Y." X 12' LIF (1) BAG GLASSINE PLAIN (1) CAUTERY TIP POLISHER LIF (1) DRAPE LAP. ABDOMINAL WITH POUCH 102" X 122" X 78" LIF (1) PK. STRIP STERI CLOSURE W' X 4" LIF (1) SKIN MARKER INK WITH 8/LABEL/TIME OUT/RULER L/F (1) CORD LAPAROSCOPY MONOPOLAR FOOD SWICHING (1) COUNTER NEEDLE BLADE 30c MAG/FOAM (1) TIME OUT BEACON NON WOVEN LIF (1) GOWN XL SMS IMPERVIOUS REINFORCED LIF (1) GOWN SURG. REINFORCED X LARGE TOWEL/WRAP LIF (1) TOWEL ABSORBENT 15" X 20" LIF (1) BLADE SURGICAL #11 CARBON STEEL (1) TUBING INSUFFLATION OPEN TIP WITHOUT VENT LIF (1) YANKAUER OPEN TIP CLEAR WITHOUT C IV LIF ATTACHED PACK WITH THE FOLLOWING STERILE COMPONENT: (1) ACTIFOG WITH/FOAM PAD SOLUTION LIF Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Status
- Terminated
- Initiated
- 2014-05-20
- Posted
- 2014-08-29
- Terminated
- 2017-08-14
- Reason
- Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
- Root cause
- Package design/selection
- Action
- Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
- Quantity
- 1 lot/60 units
- Distribution
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot, serial or model codes
- Product code 900-2270,1 lot 112125073
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FDE
- Firm FEI
- 2648727
- Firm city
- Fajardo
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28