FDA Medical Device Recalls (openFDA)
Covidien: Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, Puritan Bennett 560 Ventilator EMEA/4096600-05, Puritan Bennett 560 Ventilator CAN ANZ/4096600-06, Puritan Bennett 540 Ventilator USA/4097100 Puritan Bennett 520 Ventilator/4098300, Puritan Bennett 520 Ventilator EU-DIV/4098300-01, Puritan Bennett 520 Ventilator Japan/4098300-02, Puritan Bennett 520 Ventilator APAC/4098300-03, Puritan Bennett 520 Ventilator EMEA/4098300-05, Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06, DL4096600 DL PB560 VENTILATOR/DL4096600, DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06, REFURBISHED PB540 VENTILATOR/DL4097100, Puritan Bennett 540 Ventilator Refurbished/DS4097100
- Recall number
- Z-2446-2024
- Recalling firm
- Covidien
- Product
- Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, Puritan Bennett 560 Ventilator EMEA/4096600-05, Puritan Bennett 560 Ventilator CAN ANZ/4096600-06, Puritan Bennett 540 Ventilator USA/4097100 Puritan Bennett 520 Ventilator/4098300, Puritan Bennett 520 Ventilator EU-DIV/4098300-01, Puritan Bennett 520 Ventilator Japan/4098300-02, Puritan Bennett 520 Ventilator APAC/4098300-03, Puritan Bennett 520 Ventilator EMEA/4098300-05, Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06, DL4096600 DL PB560 VENTILATOR/DL4096600, DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06, REFURBISHED PB540 VENTILATOR/DL4097100, Puritan Bennett 540 Ventilator Refurbished/DS4097100
- Status
- Open, Classified
- Initiated
- 2024-06-24
- Posted
- 2024-08-02
- Terminated
- not on file
- Reason
- Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway
- Root cause
- Device Design
- Action
- On 6/27/24, Medtronic started correction notice distribution to physicians, health care professionals, distributors, and users who were advised of the following: 1) Any single PB500 series ventilator should only be used for the 10-year labelled service life. If a patient requires ventilatory support for a longer period of time, cumulative sequential use of more than one PB500 series ventilator should be limited to 14 years. After 14 years of cumulative use, patients should be transitioned to an alternate device. 2) When approaching 14 years of cumulative use of a PB500 series ventilator, it is recommended that patients or their caregivers discuss transitioning to an alternate ventilator with the prescribing physician. 3) Review the correction notice information with all members of your staff who need to be aware of the contents of this communication. 4) Provide the correction notice to those who need to be aware within your organization or to any organization or person where the product has been transferred or distributed. 5) Medtronic has updated the labeling. Please refer to the User Manual and Clinician s Manual addendum 6) Complete and return the response form via email to rs.gmbmitgfca@medtronic.com If you have any questions regarding this communication, please contact your Medtronic Representative or Technical Service at 1-800- 255-6774.
- Quantity
- 34,642
- Distribution
- US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Lithuania, Malaysia, Mexico, Montenegro, Namibia, Nepal, New Zealand, Nigeria, Panama, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Taiwan, Turkey, Ukraine, United Kingdom, Viet Nam, South Africa, Austria, Kenya, Latvia, Lebanon, Mongolia, Nicaragua, Thailand, Tunisia, Algeria, Jordan, Libya, Malawi, Qatar, Pakistan, United Arab Emirates, Oman, Peru, Philippines, Bhutan, Iran - Islamic Republic Of, Bolivia, Kuwait, Egypt, Switzerland, Sudan, Norway, Korea - Republic Of, Bahrain, Ethiopia, Paraguay
- Lot, serial or model codes
- REF/UDI-DI(GTIN): 4096600/10884521087798 4096600-01/10884521195998 4096600-02/10884521196001 4096600-03/10884521196018 4096600-04/10884521196025 4096600-05/10884521196032 4096600-06/10884521196377 4097100/4097100 4098300/10884521183636 4098300-01/10884521195943 4098300-02/4098300-02 4098300-03/10884521195967 4098300-05/10884521195981 4098300-06/10884521196384 DL4096600/10884521786509 DL4096600-06/10884521786516 DL4097100/DL4097100 DS4097100/DS4097100 All device serial numbers used before distribution of PB500 Series Ventilator User Manual and Clinician's Manual addendum, PT00196729 A, 2024-06-11
- 510(k) numbers
- ["K070899"]
- PMA numbers
- not on file
- Product code
- CBK
- Firm FEI
- 2024500
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28