FDA Medical Device Recalls (openFDA)
Customed, Inc: CENTRAL LINE FULL BODY DRAPE CUSTOMED CONTENTS: (1) WRAPPER 30" X 30" (2) MASK SURGICAL W/ VISOR ANTI-FOG 160mm Hg (2) CAP NURSE 21" GREEN BOUFFANT (2) TOWELS ABSORBENT 15" X 20" (2) GOWN LARGE STANDARD SMS VELCRO/NECK (1) DRAPE SHEET 41" X 58" SMS (1) DRAPE BRACHIAL ANGlO 44" X 31" REINFORCED 3" ADHESIVE CIRCULAR FENESTRATION Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Recall number
- Z-2446-2014
- Recalling firm
- Customed, Inc
- Product
- CENTRAL LINE FULL BODY DRAPE CUSTOMED CONTENTS: (1) WRAPPER 30" X 30" (2) MASK SURGICAL W/ VISOR ANTI-FOG 160mm Hg (2) CAP NURSE 21" GREEN BOUFFANT (2) TOWELS ABSORBENT 15" X 20" (2) GOWN LARGE STANDARD SMS VELCRO/NECK (1) DRAPE SHEET 41" X 58" SMS (1) DRAPE BRACHIAL ANGlO 44" X 31" REINFORCED 3" ADHESIVE CIRCULAR FENESTRATION Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Status
- Terminated
- Initiated
- 2014-05-20
- Posted
- 2014-08-29
- Terminated
- 2017-08-14
- Reason
- Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
- Root cause
- Package design/selection
- Action
- Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
- Quantity
- 8 lots/1232 units, (multiple units per lot)
- Distribution
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot, serial or model codes
- Product code 900-2262,8 lots 112041231 113036345 113047112 113078310 113088797 140111207 140211823 140312128
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 2648727
- Firm city
- Fajardo
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28