Skip to the content

FDA Medical Device Recalls (openFDA)

Clark Laboratories, Inc. (dba,Trinity Biotech USA): Trinity Biotech, Captia VZV IgG. Product code 2325600: 96 Tests; product code: 2325601: 480 tests. For in vitro diagnostic use. Intended for the detection and quantitative determination of IgG antibody to VZV in human sera.

Recall number
Z-2445-2012
Recalling firm
Clark Laboratories, Inc. (dba,Trinity Biotech USA)
Product
Trinity Biotech, Captia VZV IgG. Product code 2325600: 96 Tests; product code: 2325601: 480 tests. For in vitro diagnostic use. Intended for the detection and quantitative determination of IgG antibody to VZV in human sera.
Status
Terminated
Initiated
2012-05-29
Posted
2012-09-24
Terminated
2016-05-24
Reason
The Positive Control for the Captia Varicella-Zoster Virus (VZV) IgG Kit has been reported as running high out of its assigned ISR range.
Root cause
Nonconforming Material/Component
Action
Trinity Biotech sent a "PRODUCT RECALL LETTER" dated May 29, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Product Recall Fax Back Form was included in the letter for customers to complete and return via fax to (716) 488-1990. Customers were instructed to discard any of the affected product. Contact the Trinity Biotech Help Desk at 800-325-3424, option 2 for questions regarding this recall.
Quantity
810 kits
Distribution
Nationwide Distribution-including Puerto Rico and the states of AL, CA, IL, IN, MA, MD, NJ, OH, PA, TN, TX.
Lot, serial or model codes
Kit lot codes: 2325600-566, 2325601-566, 2325600-567, 2325601-567, 2325600-568
510(k) numbers
["K883427"]
PMA numbers
not on file
Product code
GQX
Firm FEI
1318354
Firm city
Jamestown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register