Skip to the content

FDA Medical Device Recalls (openFDA)

Vistakon: Primary (inner) Package Label: ACUVUE OASYS BRAND CONTACT LENSES for ASTIGMATISM with HYDRACLEAR Plus (senofilcon A) Johnson & Johnson Vision Care, Inc. Rx Only, STERILE. Lot B006WBS1 -9.00D -1.25 CYL 80 , 2013/05 AXIS. (Correct labeling) Mislabeled Outer (secondary) Label: Lot B006X641 D -8.50 CYL -1.25 AXIS 100, 2013/05 DO NO PAY Free Sample Not For Sale Consumer Note: This product has been provided free of charge to your Eye Care professional to aid in proper fitting. The sale of this product is prohibited. You should not have been charged money for this product. Contents: 1 lens (38% H2) in buffered saline with up to 0.005% methyl ether cellulose.

Recall number
Z-2445-2010
Recalling firm
Vistakon
Product
Primary (inner) Package Label: ACUVUE OASYS BRAND CONTACT LENSES for ASTIGMATISM with HYDRACLEAR Plus (senofilcon A) Johnson & Johnson Vision Care, Inc. Rx Only, STERILE. Lot B006WBS1 -9.00D -1.25 CYL 80 , 2013/05 AXIS. (Correct labeling) Mislabeled Outer (secondary) Label: Lot B006X641 D -8.50 CYL -1.25 AXIS 100, 2013/05 DO NO PAY Free Sample Not For Sale Consumer Note: This product has been provided free of charge to your Eye Care professional to aid in proper fitting. The sale of this product is prohibited. You should not have been charged money for this product. Contents: 1 lens (38% H2) in buffered saline with up to 0.005% methyl ether cellulose.
Status
Terminated
Initiated
2010-07-16
Posted
2010-09-21
Terminated
2011-03-28
Reason
ACUVUE OASYS Brand Contact Lenses for ASTIGMATISM may be mislabeled on the outer (secondary) packaging carton. The outer (secondary) cartons labeled as Lot B006X641 -8.50D -1.25cyl. 100 Axis may actually contain primary packages from Lot B006WBS1 -9.00D -1.25cyl. 80 Axis. The inner (primary) package labeling is correct. Product was only distributed to Eye Care Professionals for use in proper f
Root cause
Error in labeling
Action
Vistakon issued an Important Product Notification dated July 28, 2010 to customers identifying the affected product, the labeling issue, and actions to be taken by the customer. Vistakon issued a separate Important Product Notification dated July 28, 2010 to distribuitors identifying the affected product, the labeling issue, and actions to be taken by the distributor. Customers are instructed to check their inventory for affected products and return a business reply card. If replacement -8.50 lenses are needed, the customer can contact Customer Service at 1-800-874-5178. Distributors are instructed to check their inventory for affected products and return them using an enclosed return label. Replacement product will be sent by the firm. Customer Service can be contacted at 1-800-874-5178. Accounts will be instructed to go to their inventory and look for the B006X641, -8.50 lot and open the secondary carton and verify the product inside. If it is the -9.00 product they can keep the product and re-slot it in the proper place within the kit (product is ok, nothing wrong with primary package or lens inside) and order a replacement -8.50 if necessary. Business reply cards will be supplied for accounts to return to third party indicating if they had any mix/mislabeled product.
Quantity
2,123 lenses
Distribution
Worldwide Distribution: USA including all states, District of Columbia and Puerto Rico, and the countries of Canada, Bahamas, Honduras, and United Arab Emirates.
Lot, serial or model codes
Lot B006X641
510(k) numbers
not on file
PMA numbers
not on file
Product code
LPM
Firm FEI
1000222023
Firm city
Jacksonville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register