Zimmer Inc.: Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031. Surgical instrument.

Recall number
Z-2445-2008
Recalling firm
Zimmer Inc.
Product
Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031. Surgical instrument.
Status
Terminated
Initiated
2007-08-06
Posted
2008-09-20
Terminated
2009-03-11
Reason
Inadequate weld on tracker, knob oversized and poor thread design on attachment screw results in seizing. The instrument may break during use, resulting in surgical delay and an increased risk of infection.
Root cause
Device Design
Action
Firm's sales representatives were notified via e-mail on 8/6/07 to review their inventory and that of their accounts and to notify the firm of their location so that they could be replaced. Contact the firm by telephone at 574-267-6131 for additional information concerning this recall.
Quantity
39
Distribution
Nationwide Distribution including states of California, Florida and Washington.
Lot, serial or model codes
Lots 050433, BP050199, BP050258 and BP050350.
510(k) numbers
["K030827"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register