Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien LLC: Covidien Argyle Infant Heel Warmers (with tab) Item Code: MH00002T Product Usage: Intended for application of the right amount of heat to an infants heel for blood sampling.

Recall number
Z-2444-2015
Recalling firm
Covidien LLC
Product
Covidien Argyle Infant Heel Warmers (with tab) Item Code: MH00002T Product Usage: Intended for application of the right amount of heat to an infants heel for blood sampling.
Status
Terminated
Initiated
2015-07-31
Posted
2015-08-21
Terminated
2017-05-31
Reason
Infant Heel warmers may break during activation resulting in spillage of the contents on patients or healthcare worker causing potential skin/eye irritations
Root cause
Package design/selection
Action
Medtronic/Covidien issued an Urgent Medical Device recall letter dated July 31, 2015 via mailed Certified. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to quarantine and discontinue use of the affected products.. For questions or concerns,contact Medtronic representative or Customer Service at (800)-882-5878.
Quantity
164, 749 units
Distribution
Worldwide Distribution - US Nationwide and the countries: Canada, Mexico, Panama, and Japan.
Lot, serial or model codes
Lot Codes: 309907X 310604X 310704X 312110X 312329X 313034X 313615X 314212X 314308X 317921X 318904X 320714X 320715X 321312X 321423X 321712X 321812X 321826X 321903X 322120X
510(k) numbers
["K964523"]
PMA numbers
not on file
Product code
MPO
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register