Skip to the content

FDA Medical Device Recalls (openFDA)

Osypka Medical Inc: St. Jude Medical, Model 3085, Dual-Chamber External Pulse Generator The PACE 203H / Model 3085 external pulse generator / temporary cardiac pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203H / Model 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.

Recall number
Z-2444-2010
Recalling firm
Osypka Medical Inc
Product
St. Jude Medical, Model 3085, Dual-Chamber External Pulse Generator The PACE 203H / Model 3085 external pulse generator / temporary cardiac pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203H / Model 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.
Status
Terminated
Initiated
2010-06-30
Posted
2010-09-21
Terminated
2012-08-10
Reason
The recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a certain manufacturing period malfunction (i.e., not stimulate) when a type of 9 V batteries with higher than usual open circuit voltage (OCV) is used. Failure t
Root cause
Device Design
Action
OSYPKA Medical sent "Urgent Medical Device Recall" letters dated June 30, 2010 and July 7, 2010 to the direct customers. The recall notice informed the customers of the Reason for Recall, the Potential Hazard, the Temporary Fix of the Problem, the Permanent Fix of the Problem and Recommended Actions to be taken by the customer. The customers were instructed to contact Osypka Medical via telephone (858) 454-0021 or e-mail to m.osypka@osypkamed.com to verify that the Model 3085 and/or PACE 203H is subject to this product removal, obtain a return-to-manufacturer-authorization (RMA) number ,and then return the affected devices to Osypka Medical. If you have any questions call (858) 454-0021 or e-mail to mail@osypkamed.com
Quantity
12 units (US Only)
Distribution
Nationwide Distribution: USA including states of CA & FL
Lot, serial or model codes
S/N 0311030, 0323107, 0912101 & 0912110 (total = 12)
510(k) numbers
["K020896"]
PMA numbers
not on file
Product code
DTE
Firm FEI
3004185614
Firm city
La Jolla
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register