Hero Health: The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app.

Recall number
Z-2443-2021
Recalling firm
Hero Health
Product
The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app.
Status
Terminated
Initiated
2021-07-02
Posted
not on file
Terminated
2023-11-02
Reason
Due to dispenser prompting for a missed dose that had already been dispensed
Root cause
Under Investigation by firm
Action
On 07/02/2021, the firm initially sent out an "URGENT NOTICE" via email notifying customers of an product update correction. Customers were informed that the firm had become aware that a recent firmware update (v2.4.12) that users have experienced an error, which caused improper recording of a dispensed dose as having not been dispensed. This error may prompt a user to dispense the same dose again, and the mobile app will indicate that the previously dispensed dose has not been taken yet. Customers are instructed to: -If the pill dispenser prompts you to re-dispense a scheduled dose which has already been dispensed today, select Skip. -Use extra caution when dispensing doses prior to the scheduled firmware update and verify that the pill dispenser has a working internet connection. This will ensure that the firmware update (v2.4.14) will be automatically applied on all potentially affected online pill dispensers (at approximately 2:00 AM on July 3, 2021) to correct the issue. -To manually update firmware immediately, follow these steps: 1) Ensure the Pill Dispenser is connected to WiFi (Check for the WiFi icon on the main screen). 2) Go to the Device menu, by pressing the Center button on the main screen. 3) Scroll down and select Device Settings . 4) Scroll down and select Firmware Update. 5) Select I m Ready. The update could take several minutes. It is recommended to run the firmware update at least 30 minutes before a scheduled dose. -If the pill dispenser is not connected to WiFi, the firmware update cannot be completed. In this event, contact the Recalling Firm's Member support for assistance before continuing to use the pill dispenser to ensure the update is applied as soon as possible For question or further assistance, contact the Recalling firm at 855-855-9962 or via email at hello@herohealth.com (24/7).
Quantity
12,039 units which had software version V2.04.012
Distribution
U.S. Nationwide distribution.
Lot, serial or model codes
Model 100 (H-100) Catalog Number: F00002 Hero 100 V1.4
510(k) numbers
not on file
PMA numbers
not on file
Product code
NXB
Firm FEI
3014660737
Firm city
Hayward
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register