Elekta Instrument AB: Leksell Vantage Stereotactic System, UDI/GTIN 7340048304887

Recall number
Z-2442-2020
Recalling firm
Elekta Instrument AB
Product
Leksell Vantage Stereotactic System, UDI/GTIN 7340048304887
Status
Terminated
Initiated
2020-06-10
Posted
not on file
Terminated
2024-08-27
Reason
The locking mechanism at the interface of the Instrument Carrier and the Leksell Vantage Arc may not function properly.
Root cause
Device Design
Action
The firm notified their consignees on 06/10/2020 by email. The notice explained the problem and provided directions on the removal of the spring. The firm offered assistance with the removal and provided the following contact: Neurosupport@elekta.com.
Quantity
51 devices
Distribution
MA, IL, NH, NC, UT, OH, FL, MN, TX, NY, LA, AZ, MT, WV, Austria, Croatia, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Kuwait, Lithuania, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Switzerland, United Kingdom
Lot, serial or model codes
Serial numbers: SH00026, SH00049, SH00061, SH00082, SH00092, SH00111, SH00028, SH00050, SH00072, SH00084, SH00094, SH00118, SH00029, SH00052, SH00073, SH00085, SH00095, SH00121, SH00030, SH00054, SH00074, SH00086, SH00097, SH00123, SH00032, SH00055, SH00077, SH00087, SH00099, SH00127, SH00033, SH00057, SH00078, SH00088, SH00100, SH00162, SH00041, SH00058, SH00079, SH00089, SH00102, SH00044, SH00059, SH00080, SH00090, SH00105, SH00045, SH00060, SH00081, SH00091, SH00108
510(k) numbers
["K171123"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3002806883
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register