Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213

Recall number
Z-2439-2019
Recalling firm
Becton Dickinson & Co.
Product
BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213
Status
Terminated
Initiated
2019-07-26
Posted
not on file
Terminated
2019-12-06
Reason
An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
Root cause
Software design
Action
BD initiated a field correction via letter on July 26, 2019 and BD Technical Service has completed the correction (software upgrade) for all six (6) affected US customers. Letter identifies issue, health risk and customer action to take. BD is in the process of evaluating the log files of customers using the InoqulATM / InoqulA+TM software version 20.3 to determine if any users potentially experienced mismatch between a specimen and plate. BD will contact all customers using InoqulATM / InoqulA+TM software version 20.3 to provide the outcome of this evaluation. Contact: BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside the US, contact your local BD representative or distributor.
Quantity
9 units
Distribution
Illinois, Iowa, Missouri, New York, Ohio, Tennessee World Wide (5): Singapore, Taiwan, France, Netherlands
Lot, serial or model codes
Catalog Number: 447213 Serial Number/UDI: INO-000290 00382904472139  INO-000263 00382904472139 INO-000264 00382904472139 20130276 00382904472139 INO-000282 00382904472139 INO-000284 00382904472139 INO-000272 00382904472139 INO-000274 00382904472139
510(k) numbers
not on file
PMA numbers
not on file
Product code
JTC
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register