Skip to the content

FDA Medical Device Recalls (openFDA)

International Technidyne Corporation: Integrated Data Management System - Version 7.1, 7.2 and 7.3; ITC Corporation, 8 Olsen Place South, Piscataway, NJ 08854

Recall number
Z-2438-2010
Recalling firm
International Technidyne Corporation
Product
Integrated Data Management System - Version 7.1, 7.2 and 7.3; ITC Corporation, 8 Olsen Place South, Piscataway, NJ 08854
Status
Terminated
Initiated
2010-04-09
Posted
2010-09-20
Terminated
2011-05-25
Reason
A tabulation error in the Integrated Data Management Systems (IDMS) data management software that results in incorrect Electronic Quality Control (EQC) Summary Reports: The Operator and the Device Performance Summary Reports. The IDMS software is not used for the clinical management of individual patients.
Root cause
Software design
Action
ITC sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 10, 2010 to Distributors and First Consignee End Users identifying the affected product and affected summary reports, the reason for recall, and the actions to be taken by the distributors and customers. New Errata Instructions were enclosed with each letter and are to be put in the idms user guide for reference. The new instructions are to be followed when running reports for the ESV or TVT tests run by the Hemochron Response. The enclosed Customer Account Tracking Form is to be completed and returned to ITC indicating that the letter has been received and appropriate questions answered as to the use of the software system. Distributors are asked to check inventory for recalled product. If affected product is found, shipment should be stopped and placed on hold. Distributors should notify customers of the correction and provide them with a copy of the correction letter and updated instructions. This should be done by traceable mail. Distributors should return the account tracking form and return affected software disks for replacement. Questions should be directed to ITC Technical Support at 800-631-5945.
Quantity
84 total idms software systems; 66 in US, 18 outside of US.
Distribution
Worldwide distribution: USA including the states of AL, AZ, CA, FL, GA, IL, KS, LA, MD, MI, NC, ND, NJ, NM, NY, OH, PA, TN , TX, UT, VA, WA, and WI, and the countries of Italy, Turkey, Canada, France, Belgium, Germany, UAE, and New Zealand.
Lot, serial or model codes
Product code 464100, Version numbers 7.1.0, 7.2.0 and 7.3.0.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
2250033
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register