FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: (1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.

Recall number
Z-2437-2020
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
(1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.
Status
Terminated
Initiated
2020-05-15
Posted
not on file
Terminated
2021-05-18
Reason
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review
Root cause
Software design
Action
Siemens issued an Urgent Medical device Correction (UMDC ISW-20-01.A.US) and Urgent Field Safety Notification (UFSN ISW-20-01.A.OUS) to affected Customers via overnight mail on 5/15/20. The UMDC ISW 20-01.A.US and UFSN ISW-20-01. Letter states reason for recall, health risk and action to take: Manually Review QC results in CentraLink/Atellica DM daily; Review this letter with your Medical Director. If your laboratory may be affected by this issue based on the conditions listed above, you may consider the following steps to avoid this issue from occurring until Siemens updates your driver: Do not process QC while the interface communication is interrupted Use the Control lot number as the Control ID. Customers were requested to respond via the attached effectiveness check form to indicate receipt and understanding of the letter. Siemens will contact all affected customers to schedule a date to update the interface driver to version 1.5.1, which prevents the issue fromoccurring.
Quantity
186 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Bahamas, Belgium, Brazil, Chile, Colombia, Czech Republic, Egypt, Germany, Greece, Hungary, India, Italy, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Oman, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
Lot, serial or model codes
Software versions: CentraLink Data Management System software version 16.0.2 UDI: 00630414220635 CentraLink Data Management System software version 16.0.3  UDI: 00630414237268
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register