NIHON KOHDEN ORANGEMED, INC: Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.

Recall number
Z-2435-2021
Recalling firm
NIHON KOHDEN ORANGEMED, INC
Product
Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.
Status
Terminated
Initiated
2021-07-20
Posted
not on file
Terminated
2023-04-21
Reason
Potential defective gas inlet pressure regulator within ventilator which could result in gas releasing noise from the bottom of the ventilator, a High Air/O2 Inlet Pressure alarm, a High/Low O2 alarm, or could not pass device Check prior to use.
Root cause
Nonconforming Material/Component
Action
The firm Nihon Kohden OrangeMed, Inc., sent an "URGENT-Medical Device Correction Recall Safety Notice" letter to its customers on July 26, 2021. The letter describes the product, issues and actions to be taken. The customers were instructed to do the following prior to actions being implemented by the manufacturer: 1. Pass this notice on to all those who need to be aware within your organization and to any organization where the affected devices have been transferred. 2. Perform Device Check and Circuit Check as instructed in the NKV-550 ventilator Operator s Manual. A defective regulator can be identified during the Device Check and Circuit Check. 3. If the gas outlet from the hospital wall has an individually adjustable pressure regulator, adjust the outlet pressure to 16 psi for the air and O2 gases at the NKV-550 ventilator. This will allow the ventilator not to depend on its internal regulators to regulate the gas pressure. 4. If the ventilator has a failed internal regulator, remove the ventilator from service and contact the manufacturer or your local representative of the manufacturer. 5. Please complete the attached acknowledgement form and return to FSN@orange-med.com within 5 days of receipt. Please review and distribute the information with your staff members who need to be aware of the contents of this communication. Please retain a copy of this notice and include it with the Operator s Manual of this device. For further information, please contact your local representative of the manufacturer or email the manufacturer at FCN@orange-med.com. Telephone (Monday through Friday, 8:00 am to 5:00 pm, Pacific Time) Main Office: +1 (949) 502-6448 Customer Service: Ext: 7101 Technical Support: Ext. 7102 Clinical Support: Ext. 7103
Quantity
1,999 units
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, NY, MA, VT, LA, DC, PA, FL, UT and OUS (countries) of: Brazil, Japan, Ireland, Poland, Scotland, Germany, Kazakhstan, Spain, Slovenia, Germany/Bulgaria, Russia, India, Palestine, Libya, Jordan Kuwait, Lebanon, Syria, Turkey, South Africa, United Arab Emirates, Mexico, Colombia, Peru, Costa Rica, Barbados, Ecuador, Bahamas, Bolivia, St Kitts & Nevis, St. Martin, Chile, Thailand, Laos, Philippines, Malaysia, Nepal and United Kingdom.
Lot, serial or model codes
All serial numbers between NKV5502019xxxx - NKV5502112xxxx and the following serial numbers below: NKV55019270055, NKV55019370098R, NKV55019430125R, NKV55019470156R, NKV55019470160, NKV55020130200, NKV55020140213R, NKV55020140226R, NKV55020140229, NKV55020150241, NKV55020150252, NKV55020160268, NKV55020160276, NKV55020160277, NKV55020160292R, NKV55020160296, NKV55020160297, NKV55020160304, NKV55020160307, NKV55020160308, NKV55020160310, NKV55020160312, NKV55020170317, NKV55020170318, NKV55020170325, NKV55020170335, NKV55020170336, NKV55020170341, NKV55020170342, NKV55020170343, NKV55020170346, NKV55020170347, NKV55020170349, NKV55020170350, NKV55020170352, NKV55020170354, NKV55020170357, NKV55020170360, NKV55020180365, NKV55020180366, NKV55020180369, NKV55020180370, NKV55020180371, NKV55020180372, NKV55020180373, NKV55020180374, NKV55020180375, NKV55020180376, NKV55020180378, NKV55020180379, NKV55020180381, NKV55020180383, NKV55020180384, NKV55020180385, NKV55020180387, NKV55020180391, NKV55020180392, NKV55020180396, NKV55020180397, NKV55020180398, NKV55020180399, NKV55020180400, NKV55020180401, NKV55020180402, NKV55020180403, NKV55020180405, NKV55020180406, NKV55020180408, NKV55020180409, NKV55020180410, NKV55020180411, NKV55020180412, NKV55020180413
510(k) numbers
["K181695","K192307"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3014631252
Firm city
Santa Ana
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register