FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332
- Recall number
- Z-2435-2019
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332
- Status
- Terminated
- Initiated
- 2019-08-07
- Posted
- not on file
- Terminated
- 2020-06-24
- Reason
- A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instrument and resulting in premature need for replacement and/or breakage of the instrument
- Root cause
- Nonconforming Material/Component
- Action
- Integra LifeSciences notified consignees (Hospitals, Distributors) by letter delivered via FedEx on August 7, 2019. Letter states reason for recall, health risk and action to take: 1. Review your inventory and determine if you have any impacted lots listed above.2. If you do have product with an affected Lot Number, stop using the product immediately and remove the product from service. 3. Complete the attached form. If you do not have affected product, check the box, I do not have any product with the lot number. 4. If you do have affected product, check the box I do have product(s) with the lot number. Record the total quantity of affected product you have.5. Complete the other information and Return the completed Acknowledgement and Return Form by email or fax indicating your receipt and review of this notification. Keep a copy of the form for your records. Questions contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com Outside USA: 781-565-1401.
- Quantity
- 237 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Lot #AA1901
- 510(k) numbers
- ["K102048"]
- PMA numbers
- not on file
- Product code
- DXC
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28