Skip to the content

FDA Medical Device Recalls (openFDA)

Arjo, Inc.: Kwiktrak Gate System; the rail system used with the Maxi Sky 600 Ceiling Patient Lift; The Kwiktrak Gate System consists of part 700.11560 - Kwiktrak gate motor box (serialized), part 700.11550 - Kwiktrak gate fixed box, part 700.11505 - Kwiktrak gate combination kit (includes 2 x 700.11550 and 2 x 700.11560), and part 700.11500 - Kwiktrak gate (includes 1 x 700.11550 and 1 x 700.11560) Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. The equipment is intended for transferring lifts from one system track to another.

Recall number
Z-2434-2008
Recalling firm
Arjo, Inc.
Product
Kwiktrak Gate System; the rail system used with the Maxi Sky 600 Ceiling Patient Lift; The Kwiktrak Gate System consists of part 700.11560 - Kwiktrak gate motor box (serialized), part 700.11550 - Kwiktrak gate fixed box, part 700.11505 - Kwiktrak gate combination kit (includes 2 x 700.11550 and 2 x 700.11560), and part 700.11500 - Kwiktrak gate (includes 1 x 700.11550 and 1 x 700.11560) Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. The equipment is intended for transferring lifts from one system track to another.
Status
Terminated
Initiated
2008-08-01
Posted
2008-09-20
Terminated
2010-01-05
Reason
There is a possibility that the Kwiktrak Gate System could malfunction, resulting in the gate locks opening even though the corresponding tracks are not properly aligned.
Root cause
Device Design
Action
Arjo Inc. USA sent Urgent Field Safety Notice letters dated 8/1/08 to the direct accounts on 8/1/08, informing them that there is a remote possibility that the gate system could malfunction. The result of this malfunction is that the gate locks could open even though the corresponding tracks are not properly aligned. This could result in the lift/cassette falling off the track system, possibly resulting in serious personal injuries and or property damage. The letters requested the customer to immediately implement special operating instructions in their facility to ensure great care is used when the gate system is operated. All personnel must be made aware that the tracks need to be properly aligned before using the gate. Illustrations of an unaligned gate system with an open lock and properly aligned tracks and gate system with open locks were included. Any questions the customers may have were directed to the Arjo Quality Department at 1-800-323-1245, ext. 6125. The accounts were informed that Arjo personnel will contact them in September 2008 to install a permanent add-on improvement to prevent the occurrence of a malfunction. The accounts were requested complete the enclosed response form indicating that they have read and understood the 8/1/08 letter, and fax it back to Arjo at 888-594-2756.
Quantity
230 units
Distribution
Nationwide Distribution --- including states of Arizona, Colorado, Illinois, Indiana, Michigan, Minnesota, Missouri, Montana, New York, North Carolina, Ohio, Oklahoma, Texas, and Wisconsin.
Lot, serial or model codes
part 700.11560, serial numbers GA-1304-0001 to GA-2608-1374, and part 700.11550, all units
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
1419652
Firm city
Roselle
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28