Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries Inc: Medline Electrical Clipper with Charging base The Electrical Surgical Clipper is intended for the removal of body hair from patients as required prior to surgery or other minor procedures.

Recall number
Z-2433-2015
Recalling firm
Medline Industries Inc
Product
Medline Electrical Clipper with Charging base The Electrical Surgical Clipper is intended for the removal of body hair from patients as required prior to surgery or other minor procedures.
Status
Terminated
Initiated
2015-07-08
Posted
2015-08-20
Terminated
2017-11-29
Reason
Charging base of surgical clippers overheats and smokes.
Root cause
Process control
Action
Medline Industries sent a " Urgent Remedial Action Destruction Form " letter via first class dated July 9, 2015, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customer. ***URGENT RECALL IMMEDIATE ACTION REQUIRED SURGICAL CLIPPER CHARGER BASE DYND70800 *** Medline Industries, Inc. is recalling specific lots of the Medline Surgical Clippers Charger Base (base only). This recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base malfunction. These lots were distributed October 2013 through March 2015. After further investigation we have determined the root cause to be a defective circuit board. This defect has since been corrected. REQUIRED ACTION: 1. Immediately check your stock for item number DYND70800 and the affected lots listed on the destruction form, dispose of the affected charger base only (do not dispose of the clipper). 2. Please return the completed enclosed destruction form listing the quantity of affected charger base(s) you destroyed. Even if you do not have any affected product please complete and return the form, as indicated on the destruction form. When we receive your completed destruction form we will send you replacement charger base(s). 3. If you are a distributor, promptly notify any of your customers that may have received affected product. If you have any questions, please contact 866-359-1704. As the distributor of this product, we sincerely regret the inconvenience. We, like you, place the health and safety of your patients first and foremost. This recall is being made with the knowledge of the U.S. Food and Drug Administration.
Quantity
7062 units (U.S.); 540 units Foreign
Distribution
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and to the countries of : United Arab Emirates, Barbados, Bahamas, Chile, Costa Rica, Italy, Lebanon, and Singapore.
Lot, serial or model codes
Product item number: DYND70800 with Lot Numbers: 13F1, 13KD, 14A1, 14C1, and Lot 1.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LWK
Firm FEI
1417592
Firm city
Mundelein
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register