Siemens Healthcare Diagnostics Inc: RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778

Recall number
Z-2432-2020
Recalling firm
Siemens Healthcare Diagnostics Inc
Product
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Status
Terminated
Initiated
2020-05-27
Posted
not on file
Terminated
2021-01-12
Reason
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
Root cause
Software design
Action
On May 27, 2020, the firm notified customers via Urgent Field Safety Notice letters. Product was distributed to foreign customers only. Customers were instructed to complete and return the included customer response form. Customers whose systems are not configured to use code 39 with check-digit need take no further action. However, for customers whose systems are configured to use code 39 with check-digit, the customer notice provided two workaround options: 1) Use the integrated on-board barcode scanner, or 2) Re-configure the handheld barcode scanner following the instructions in the letter. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative
Quantity
436
Distribution
Foreign distribution only.
Lot, serial or model codes
All units
510(k) numbers
["K192240"]
PMA numbers
not on file
Product code
CHL
Firm FEI
3002637618
Firm city
Norwood
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register