Skip to the content

FDA Medical Device Recalls (openFDA)

Remel, Inc: Xpect Giardia/Cryptosporidium Kit, Ref 24050520, 20 tests per kit, For In Vitro Diagnostic Use. A rapid test for the direct qualitative detection of Giardia and Cryptosporidium antigens in fecal specimens.

Recall number
Z-2432-2008
Recalling firm
Remel, Inc
Product
Xpect Giardia/Cryptosporidium Kit, Ref 24050520, 20 tests per kit, For In Vitro Diagnostic Use. A rapid test for the direct qualitative detection of Giardia and Cryptosporidium antigens in fecal specimens.
Status
Terminated
Initiated
2008-08-11
Posted
2008-09-19
Terminated
2010-01-21
Reason
Faint grey test lines for Cryptosporidium may be visible which may be interpreted as false positive
Root cause
Process control
Action
The recalling firm issued letters dated 6/19/08 flagged "IMPORTANT PRODUCT INFORMATION" which advised their customers of the issue with the faint grey test lines but did not request any action be taken, such as return or destruction of the affected lot numbers. The recalling firm issued a second letter flagged "IMPORTANT MEDICAL DEVICE CUSTOMER ADVISORY NOTICE" dated 8/11/08 and was issued via regular mail on 8/14/08. This letter asked users to review their inventory for the product lots indicated in the letter, discard all remaining units of this product lot, complete and return the Product Inventory Checklist regardless if they have affected product. A "PRODUCT INVENTORY CHECKLIST" was enclosed to report whether the firm had product to discard and the amount to be replaced.
Quantity
3,974 kits
Distribution
Worldwide Distribution --- including USA and countries of Canada and U.K.
Lot, serial or model codes
Lot numbers 597374, Exp. 2008.10.08; 607636, Exp. 2008.11.06; 615970, Exp. 2008.12.02; 615971, Exp. 2008.12.02; 621187, Exp. 2008.12.02; 622617, Exp. 2009.04.02; 624618, Exp. 2009.01.02; 635553, Exp. 2009.04.17; 635554, Exp. 2009.04.24; 640609, Exp. 2009.05.01; 654829, Exp. 2009.05.30; and 658878, Exp. 2009.05.30
510(k) numbers
["K031834"]
PMA numbers
not on file
Product code
MHI
Firm FEI
1924669
Firm city
Lenexa
Firm state
KS
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register