Skip to the content

FDA Medical Device Recalls (openFDA)

Sekisui Diagnostics Llc: Sekisui Spectrolyse PAI-1, In-Vitro Diagnostic for Quantitative determination of Plasminogen Activator Inhibitor Type- 1 (PAI-1) activity in human plasma. Catalog Number: 101201 For the determination of Plasminogen Activator Inhibitor Type 1 (PAI-1) activity in human plasma. The test is for in Citro diagnostic use.

Recall number
Z-2431-2012
Recalling firm
Sekisui Diagnostics Llc
Product
Sekisui Spectrolyse PAI-1, In-Vitro Diagnostic for Quantitative determination of Plasminogen Activator Inhibitor Type- 1 (PAI-1) activity in human plasma. Catalog Number: 101201 For the determination of Plasminogen Activator Inhibitor Type 1 (PAI-1) activity in human plasma. The test is for in Citro diagnostic use.
Status
Terminated
Initiated
2012-08-28
Posted
2012-09-20
Terminated
2016-07-25
Reason
SPECTROLYSE PAI-1 activity assay producing lower than expected absorbance values with "0" standard. This will reduce the slope of the standard curve requiring the user to make more frequent sample dilutions.
Root cause
Process control
Action
The firm, Sekisui Diagnostics, LLC, sent an "URGENT PRODUCTS RECALL NOTICE" dated August 28, 2012 to its customers. The notice describes the product, problem and actions to be taken. The customers were instructed to immediately stop using the product; to discard any remaining inventory and complete and return the Customer Acknowledgement Form by September 4, 2012 documenting the # of kits destroyed via FAX to Quality Assurance at 203-602-5553. The customers were also instructed to return the form even if they do not have the affected product currently in inventory. Sekisui will replace kits. If you have any questions contact, Customer Service Department at 800-637-3375, ext 10, 12, or 31, or email to recall coordinator at MKHAN@AMDIAG.COM.
Quantity
699 kits
Distribution
Nationwide distribution: USA including states of: FL, IN, IL, and WA.
Lot, serial or model codes
Lot Numbers: 120425 and 120601
510(k) numbers
["K063323"]
PMA numbers
not on file
Product code
GGP
Firm FEI
1220602
Firm city
Stamford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register