Bio-Rad Laboratories: Kallestad (TM) Mouse Kidney. Product labeled "BIO-RAD Kallestad (TM) Mouse Kidney Substrate... UNITED STATES, Bio-Rad Laboratories, Redmond, WA 98052... FRANCE, Bio-Rad Laboratories, 92430 Marnes La Coquette". Catalog numbers: 30445 - 8 wells, 48 tests and 30447 - 8 well, 192 tests.
- Recall number
- Z-2430-2008
- Recalling firm
- Bio-Rad Laboratories
- Product
- Kallestad (TM) Mouse Kidney. Product labeled "BIO-RAD Kallestad (TM) Mouse Kidney Substrate... UNITED STATES, Bio-Rad Laboratories, Redmond, WA 98052... FRANCE, Bio-Rad Laboratories, 92430 Marnes La Coquette". Catalog numbers: 30445 - 8 wells, 48 tests and 30447 - 8 well, 192 tests.
- Status
- Terminated
- Initiated
- 2008-07-29
- Posted
- 2008-09-19
- Terminated
- 2011-03-31
- Reason
- The Homogeneous Positive control was released with a titer of 1:256 rather than the titer value of 1:128 reported on the QC Report Label included with each test kit.
- Root cause
- Employee error
- Action
- On 7/25/08, the firm began to notify their customers about the recall via an Urgent Product Correction sent by DHL or FEDEX. Included with the customer letter was a revised QC Card. International subsidiaries were notified by e-mail with the customer letter and revised QC Card attached. The letter informed customers of the recall and directing them to the revised QC card which was attached to the letter. The instructions are to, "Please use this to replace the card in any kits remaining in your facility." If customers have any questions or require additional QC cards, they are to contact Technical Support.
- Quantity
- 69 total kits
- Distribution
- Worldwide Distribution --- including USA and countries of AUSTRALIA, BRAZIL, CANADA, CHINA , FRANCE, GERMANY, HONG KONG, INDIA, ITALY, JAPAN, MEXICO, NEW ZEALAND, POLAND, SINGAPORE, SOUTH KOREA , TAIWAN, UNITED KINGDOM, and URUGUAY.
- Lot, serial or model codes
- Lot # and exp: for catalog # 30445 - 951563, exp 11/30/08; 953560, exp 1/31/09; and 955471, exp 4/30/09; for catalog # 30447 - 951565, exp 11/30/08; 953561, exp 1/31/09; and 954607, exp 3/31/09.
- 510(k) numbers
- ["K792610"]
- PMA numbers
- not on file
- Product code
- DBL
- Firm FEI
- 3008272466
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28