FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Radio-Capitellum, Capitellum, Left, Large, Do Not Reuse, Rx Only. Indicated for use in the elbow for reduction or relief of pain and/or improved elbow function in skeletally mature patients.
- Recall number
- Z-2428-2015
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Radio-Capitellum, Capitellum, Left, Large, Do Not Reuse, Rx Only. Indicated for use in the elbow for reduction or relief of pain and/or improved elbow function in skeletally mature patients.
- Status
- Terminated
- Initiated
- 2015-06-24
- Posted
- 2015-08-20
- Terminated
- 2016-04-29
- Reason
- Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.
- Root cause
- Packaging change control
- Action
- Branches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.
- Quantity
- 16992 total
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- Part Number: 410-0000 (Radio Capitellum Large, Left) - Lot #'s 18832, 22193201, 22193202, 22193203, 22258301,22957101, 23853901, 19069-S2, 19077-S2 and 19155S2; Part Number: 410-0001 (Radio Capitellum Small, Left) - Lot #'s18833, 18834, 22193302, 22193303, 22766801, 2295601 and 23853101; Part Number: 410-0002 (Radio Capitellum Large, Right) - Lot #'s18835, 18836, 22193401, 22193402, 22193403, 22766901, 22956301, 23853201 and 19074-S2; Part Number: 410-0003 (Radio Capitellum Small, Right) - Lot #'s 18565, 22193501, 22193502, 22193503, 22767001, 22956401 and 23853301
- 510(k) numbers
- ["K070236"]
- PMA numbers
- not on file
- Product code
- JDB
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28