Philips North America LLC: Philips, HEARTSTART, INFANT/CHILD - SMART PADS CARTRIDGE, REF: M5072A, (0-8 YEARS)< 55 lbs / 25 kg, CE 0123, NON-STERILE, Rx only, UDI: (01) 00884838023758 for use with defibrillator

Recall number
Z-2427-2021
Recalling firm
Philips North America LLC
Product
Philips, HEARTSTART, INFANT/CHILD - SMART PADS CARTRIDGE, REF: M5072A, (0-8 YEARS)< 55 lbs / 25 kg, CE 0123, NON-STERILE, Rx only, UDI: (01) 00884838023758 for use with defibrillator
Status
Terminated
Initiated
2021-07-16
Posted
2021-09-03
Terminated
2023-05-12
Reason
Infant/Child defibrillator pads contain incorrect labeling. This could lead to a delay in therapy, or delivery of lower energy than indicated for an adult patient
Root cause
Process control
Action
On July 26, 2021, Phillips issued an "Urgent Medical Device Removal" notification to affected consignees. On August 23, 2021, Phillips issued a revised "Urgent Medical Device Removal notification to update their recall strategy concerning replacement infant/child smart pads cartridges. In addition to informing consignees about the recalled product, Phillips asked consignees to take the following actions: Action 1: If needed, contact your distributor to request a replacement. Customers who purchased their pads cartridges directly from Philips will receive their replacements without any further action on their part. Customers who purchases their pads cartridges from a distributor should contact that distributor to request replacement. Replacement pads cartridges may not be available immediately, therefore, refer to Action 2, below. Action 2: Determine if the Infant/Child SMART Pads may remain in service until replaced. 1. Philips, or your Philips Distributor, will supply a replacement of all M5072A Infant/Child SMART Pads Cartridges from Lot Number Y111220-01. 2. If the Infant/Child SMART Pads Cartridge is still inside the pouch, then it can remain in service until replaced. While waiting to receive the replacement, customers should keep the Infant/Child SMART Pads Cartridge inside the outer pouch until the cartridge is replaced, or until it is needed for therapy. The reason is because the pouch correctly identifies the cartridge as intended for infant/child patients. (Figure 7). 3. If the Infant/Child SMART Pads Cartridge has been removed from the outer pouch, and if the protective cover shows a drawing of pads placement for a child s chest and back, then it can remain in service until replaced. The reason is because the protective cover has the correct image with the image of a child s chest and back. (Figure 8). 4. If the Infant/Child SMART Pads Cartridge has been removed from the outer pouch, and if the protective cover shows a dra
Quantity
19,840 units
Distribution
Worldwide distribution. US Nationwide, Australia, Austria, Bolivia, Canada, Chile, China, Denmark, France, French Polynesia, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Republic of Korea, Malta, Martinique, Mexico, Monaco, Netherlands, Norway, Poland, Spain, Switzerland, Taiwan, United Kingdom, and Uruguay
Lot, serial or model codes
Lot Y111220-01
510(k) numbers
not on file
PMA numbers
["P160029"]
Product code
MKJ
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register