FDA Medical Device Recalls (openFDA)
Smith & Nephew Inc: 12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.
- Recall number
- Z-2427-2012
- Recalling firm
- Smith & Nephew Inc
- Product
- 12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.
- Status
- Terminated
- Initiated
- 2012-08-22
- Posted
- 2012-09-20
- Terminated
- 2013-02-27
- Reason
- Modular necks were manufactured out of specification by a vendor; the taper provide is undersized.
- Root cause
- Employee error
- Action
- Smith & Nephew sent an "URGENT:CLASS II RECALL-HIGH OFFSET MODULAR NECK R1213 letter dated August 22, 2012 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Customers were instructed to cease use and distribution of the product and quarantine for return. An Inventory Return Certification Form was included for customers to complete and return via fax to 901-566-7975.
- Quantity
- 70 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CA, FL, GA, IL, IN, MA, NC, NJ, NY, OH, OK, PA, TN, TX, and WA, and the country of Canada.
- Lot, serial or model codes
- Lot Numbers: 12FTM0017A, 12FTM0017B, 12FTM0017C, 12FTM0019A. 12FTM0019B, 12FTM0019C, 12FTM0019D, 12FTM0020A, 12FTM0020B, 12FTM0020C, 12FTM0021A, 12FTM0021B, and 12FTM0021C
- 510(k) numbers
- ["K081124"]
- PMA numbers
- not on file
- Product code
- MEH
- Firm FEI
- 1020279
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28