Siemens Healthcare Diagnostics, Inc: Atellica IM BR 27.29 (BR) Assay 250 Test Kit (Reference)- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer.- SMN: 11206236

Recall number
Z-2426-2021
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
Atellica IM BR 27.29 (BR) Assay 250 Test Kit (Reference)- in vitro diagnostic use in the quantitative serial determination of cancer antigen CA 27.29 in human serum and plasma (EDTA) using the Atellica IM Analyzer.- SMN: 11206236
Status
Open, Classified
Initiated
2021-07-06
Posted
not on file
Terminated
not on file
Reason
Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot numbers
Root cause
Under Investigation by firm
Action
Siemens Healthcare issued Urgent Medical Device Correction (UMDC) via FedEx beginning on 07/06/2021. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 07/06/2021. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director. Change the BR assay Pack Calibration Interval on each Atellica IM analyzer from 10 days to 3 days. See Additional Information for instructions on changing the Pack Calibration Interval. All Atellica IM BR assay kit lots can continue to be utilized with the updated Pack Calibration Interval of 3 days. Continue to follow the Atellica IM BR IFU claimed 28-day On-board Stability and 32-day Lot Calibration Intervals. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Additional Information Instructions for Changing the Pack Calibration Interval on Atellica IM Analyzers Instructions for updating the BR Pack Calibration Interval from 10 days to 3 days on each Atellica IM analyzer are outlined below. 1) Sign into Atellica IM analyzer as Labmanager user level 2) In the Worklist Overview screen, move all BR results to Historical 3) In the Setup screen, select Test Definition, select IM Test Definition, and select BR Test 4) Select the Calibration Tab 5) Under Pack C
Quantity
787 units
Distribution
Worldwide distribution. US Nationwide, Australia, Bahamas, India, Mexico, Saudi Arabia, and Viet Nam
Lot, serial or model codes
All lots impacted Kit lot numbers: 41331255 & 74682255 expiration date 2021-08-03, Lot number: 91605257 expiration date 2021-11-10.
510(k) numbers
["K982680"]
PMA numbers
not on file
Product code
MOI
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register