Ion Beam Applications S.A.: Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
- Recall number
- Z-2424-2018
- Recalling firm
- Ion Beam Applications S.A.
- Product
- Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
- Status
- Terminated
- Initiated
- 2018-05-16
- Posted
- 2018-06-20
- Terminated
- 2018-10-16
- Reason
- Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
- Root cause
- Software design
- Action
- On May 18, 2018 IBA issued URGENT FIELD SAFETY NOTICES to their customers. Customers were advised to take the following actions: It is recommended to ensure that the correction vector confirmation message appears on the expected access point. When the correction vector is applied (if deemed necessary1), ensure that the Patient Positioning System GoTo the corrected position for treatment before starting irradiation: - verify on adaPTdeliver screen that the corrected target position is selected before engaging the movement. - verify on adaPTdeliver screen that the corrected target position appears on the screen after the movement. Moreover, as advised in the Proton Therapy System Clinical User Guide, it is recommended to verify the patient position after each correction vector application. A software update will be deployed on all affected sites by July 2018. Customers with questions may email: vigiliance@iba-group.com
- Quantity
- 3 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands
- Lot, serial or model codes
- Serial Numbers: SAT.122 (EU) SAT.132 (EU) SAT.133 (US)
- 510(k) numbers
- ["K101508"]
- PMA numbers
- not on file
- Product code
- LHN
- Firm FEI
- 3000256071
- Firm city
- Louvain La Neuve
- Firm state
- not on file
- Firm country
- Belgium
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28