FDA Medical Device Recalls (openFDA)
International Biophysics Corp.: IBC Aortic Punch Catalog Numbers APM40, APL48A, APM44A, APL40A, APM40A, CVAPL36, CVAPL40, CVAPL44, CVAPL48, CVAPL52, APM28BM, APM36BM, APM40BM, APM44BM, APM48BM, APM40BK, APM48BK, APM36BK, APM44BK, APM28, APM36, APL44, APM40 Product Usage: The IBC Aortic Punch is a single-use, disposable surgical punch used to create an opening in the wall of the aorta to prepare a site for anastomosis.
- Recall number
- Z-2423-2015
- Recalling firm
- International Biophysics Corp.
- Product
- IBC Aortic Punch Catalog Numbers APM40, APL48A, APM44A, APL40A, APM40A, CVAPL36, CVAPL40, CVAPL44, CVAPL48, CVAPL52, APM28BM, APM36BM, APM40BM, APM44BM, APM48BM, APM40BK, APM48BK, APM36BK, APM44BK, APM28, APM36, APL44, APM40 Product Usage: The IBC Aortic Punch is a single-use, disposable surgical punch used to create an opening in the wall of the aorta to prepare a site for anastomosis.
- Status
- Terminated
- Initiated
- 2015-08-07
- Posted
- 2015-08-19
- Terminated
- 2016-03-01
- Reason
- It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria.
- Root cause
- Packaging process control
- Action
- The recalling firm began notifying their affacted customers on 7/2/2015
- Quantity
- 10,330 units
- Distribution
- Worldwide - United States Nationwide and the countries of: Germany, Norway, Turkey, Chile, Pakistan, and Republic of South Africa
- Lot, serial or model codes
- 042415-3488, 022515-3372, 02215-3374, 022515-3372, 022515-3371, 022515-3373, 022515-3382, 022515-3383, 022515-3384, 022515-3385, 022515-3386, 040615-3452, 040615-3453, 040615-3449, 040615-3450, 040615-3451, 040615-3448, 040615-3447, 050815-3510, 050815-3509, 050815-3511, 050815-3512, 052715-3602, 060315-3603, 031215-3412, 022515-3387, 022515-3388, 033015-3442, 042415-3488
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- DWS
- Firm FEI
- 1000117854
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28