Skip to the content

FDA Medical Device Recalls (openFDA)

Carefusion 2200 Inc: Snowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 60 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7046; endoscopic surgical device

Recall number
Z-2422-2023
Recalling firm
Carefusion 2200 Inc
Product
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 60 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7046; endoscopic surgical device
Status
Open, Classified
Initiated
2023-07-13
Posted
2024-09-30
Terminated
not on file
Reason
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Root cause
Nonconforming Material/Component
Action
BD issued an URGENT: Medical Device Product Recall notice to its consignees on 13 July 2023 via FedEx 2-day and email. The notice explained the problem with the device, the risk to the patient, and requested the following actions: Medical facilities: - Discontinue the use of the affected device and destroy it - Share the notice with all users within your facility network - If the product was purchased through a distributor, contact the distributor directly for further instructions. Distributors were directed to Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this recall notification on BD s behalf. Questions: Phone: 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted, M-F 8am - 5pm CT Email: productcomplaints@bd.com
Quantity
4 units
Distribution
US and Canada
Lot, serial or model codes
UDI/DI 10885403160578, Lot Numbers: C23, L22
510(k) numbers
["K930667"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
3008002452
Firm city
Vernon Hills
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register