Philips Medical Systems North America Co. Phillips: DigitalDiagnost X-ray System is a stationary Diagnostic X-ray using Digital Solid State Imaging Detector. It is identified with Part Number 4512 201 04752 Software Version 1.5.1 and Part Number 4512 201 04753 Software Version 1.5.2. The device is manufactured by Philips Medical Systems DMC GmbH, RontgenstraBe 24, 22335 Hamburg, Germany.

Recall number
Z-2422-2010
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
DigitalDiagnost X-ray System is a stationary Diagnostic X-ray using Digital Solid State Imaging Detector. It is identified with Part Number 4512 201 04752 Software Version 1.5.1 and Part Number 4512 201 04753 Software Version 1.5.2. The device is manufactured by Philips Medical Systems DMC GmbH, RontgenstraBe 24, 22335 Hamburg, Germany.
Status
Terminated
Initiated
2007-12-13
Posted
2010-09-14
Terminated
2010-09-14
Reason
When using DICOM print functionality, the printed images may contain data for the wrong patient.
Root cause
Software design
Action
Philips Healthcare issued an "URGENT PRODUCT CORRECTION " letter dated March 10, 2008 to their consignees. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: 1) Validate that image and patient information are correct on hard copy prints. 2) Reprint images that have mixed patient information, and re-check that image and patient information are correct. 3) Do not use printed images if the information is incorrect. The firm will issue the Field Change Order (FCO) 71200025 and the field service engineers will visit each consignee and conduct a software upgrade to solves the printing problem. The consignees may contact the Philips Call Center at 1-800-772-9377, #5 #2 and reference "FCO 71200025".
Quantity
119 units
Distribution
Nationwide Distribution: including the states of AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VT, WA, and WI.
Lot, serial or model codes
119 units are identified with Site Numbers:  59667, 82641, 82681, 86222, 86987, 86988, 103893, 103894, 104465, 104937, 104938, 105367, 504989, 505545, 506776, 532141, 536043, 536913, 539410, 540477, 540996, 541079, 541086, 541856, 541857, 542605, 542606, 543430, 544119, 544135, 544487, 544632, 544634, 544658, 545606, 547799, 548185, 548186, 548245, 548391, 548392, 548670, 549294, 549418, 549428, 549651, 549876, 549877, 549912, 550016, 550021, 550071, 550229, 550262, 550271, 550343, 550352, 550357, 550359, 550361, 550362, 550363, 550414, 550415, 550417, 550418, 550547, 550828, 550846, 550874, 550878, 550891, 550937, 551006, 551022, 551574, 551668, 551673, 552375, 552378, 552380, 552404, 552457, 552641, 553014, 553272, 553273, 553421, 553424, 553573, 553607, 554063, 554335, 554354, 554359, 554422, 554449, 554517, 554711, 554714, 554771, 555163, 555164, 555534, 555685, 555754, 553017, 531994, 535549, 543218, 543847, 547216, 549742, 550312, 551522, 552507, 552876, 552914, and 555119.
510(k) numbers
["K982795"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register