Stryker Instruments Div. of Stryker Corporation: Stryker Bur Guard REF 2296-301, Stryker Instruments, Kalamazoo, MI (USA) The Impaction Bur Guard is intended to be used with the Impaction Drill. The drill is used for surgical procedures involving precision drilling and cutting of bone and hard tissue. The bur guard covers the nose of the handpiece and the shaft of the cutting accessory to prevent exposure to heat and moving parts, as well as minimize bur whip and handpiece vibration.

Recall number
Z-2420-2010
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Stryker Bur Guard REF 2296-301, Stryker Instruments, Kalamazoo, MI (USA) The Impaction Bur Guard is intended to be used with the Impaction Drill. The drill is used for surgical procedures involving precision drilling and cutting of bone and hard tissue. The bur guard covers the nose of the handpiece and the shaft of the cutting accessory to prevent exposure to heat and moving parts, as well as minimize bur whip and handpiece vibration.
Status
Terminated
Initiated
2010-06-02
Posted
2010-09-13
Terminated
2011-04-21
Reason
The reason for the field action is that the inner diameter of the Bur Guards may be out of specification. If there is excess material on the inside of the Bur Guard at the location where the guard tapers, it may come into contact with the Nose Cone of the drill. This could result in an injury or burn to the patient and/or user.
Root cause
Device Design
Action
Stryker Instruments sent an Urgent Medical Device Recall Notification dated June 10, 2010, to their customers. The letter described the product the problem and the action to be taken by the customers. The customers were instructed to: 1) immediately review this recall notification. 2) forward this letter and the attached Business Reply Form (BRF) to all affected locations. 3) Immediately check and quarantine any affected product found. 4) Destroy all affected product by breaking off at least 3 of the tabs on the bur guard and then dispose of the affected product in normal trash. 5) Complete and sign the BRF and fax to Stryker Instruments Regulatory Department, 866-521-2762. NOTE: Upon receipt of the BRF, a replacement bur guard will be sent to the customer's account for each affected bur guard destroyed. For questions regarding this recall please contact Stryker Instruments: 1-800-800-4236, ext. 3808 jennifer.mars@stryker.com
Quantity
10982
Distribution
Worldwide Distribution: US AND COUNTRIES OF SCANDINAVIA, AUSTRALIA, CANADA, CHILE, CHINA, BRAZIL, FRANCE, GERMANY, GREECE, SOUTH AFRICA, ITALY, JAPAN, MEXICO, SWITZERLAND, PORTUGAL, SINGAPORE and UK.
Lot, serial or model codes
Replace by dates of May 2009 through August 2010
510(k) numbers
["K961970"]
PMA numbers
not on file
Product code
HWE
Firm FEI
1811755
Firm city
Kalamazoo
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register