Boston Scientific Corporation: Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY 2011-05, REF 39014, UPN H749390140, Sterile R Sterilized using irradiation, Made in USA 47900 Bayside Parkway, Fremont, CA 94538 The intended use as stated in Japan Shonin 21200BZY00484000: This product is an imaging catheter with built-in ultrasonic transducer that sends ultrasound waves inside the blood vessel to perform B-mode ultrasonagraphy. The indications for use as stated in the 510(k) and DFU 90606867: The Atlantis coronary catheters are intended for ultrasound examination of coronary intravascular pathology only. Intravascular Ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.

Recall number
Z-2419-2011
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY 2011-05, REF 39014, UPN H749390140, Sterile R Sterilized using irradiation, Made in USA 47900 Bayside Parkway, Fremont, CA 94538 The intended use as stated in Japan Shonin 21200BZY00484000: This product is an imaging catheter with built-in ultrasonic transducer that sends ultrasound waves inside the blood vessel to perform B-mode ultrasonagraphy. The indications for use as stated in the 510(k) and DFU 90606867: The Atlantis coronary catheters are intended for ultrasound examination of coronary intravascular pathology only. Intravascular Ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Status
Terminated
Initiated
2011-03-28
Posted
2011-06-10
Terminated
2012-07-30
Reason
Catheter tip detachment that is not associated with elongation of the catheter typically caused by excessive pulling forces. EXPANDED: Boston Scientific is initiating a recall of all iCross Coronary Imaging Catheters because they have received and confirmed 8 occurrences of catheter tip detachments due to embrittlement.
Root cause
Process design
Action
Boston Scientific " Important Customer Notification" letters dated March 14, 2011 were hand delivered to customers by Boston Scientific sales representatives starting on March 28th, 2011. The sales representatives verbally reviewed information. A tracking log will be maintained document completion of the customer notifications. The letter described the product, problem and possible root cause as well as one corrective action, a change in sterilization method from Gamma to E-beam implementation. Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter dated May 26, 2011, to all affected customers. Customers were instructed to Identify and discontinue use of any affected product within their inventory and to segregate it immediately and return it to Boston Scientific
Quantity
30,669
Distribution
Worldwide Distribution - USA (nationwide) and Puerto Rico and the countries of Japan, Trinidad, Tobago, and American Virgin Islands.
Lot, serial or model codes
All lots are impacted
510(k) numbers
["K050577","K063312"]
PMA numbers
not on file
Product code
DQO
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register